5890 Series II GC with EPC, FID
$7,500*
Includes
single Split/Splitless Injector with Electronic
Pressure Control, Flame Ionization Detector, INET
Communications installed on GC, cables, manuals and ONE
(1) YEAR WARRANTY. Other injectors and detectors
available. HPIB Communications and Data System available.
Available
in HP Parts.
*All
prices are in U.S. dollars and do not include shipping, handling
and installation. International
Cooperating Organizations
|
|
|
Specifications:
Heated Devices
- Five heated
zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored:
Two
Split/Splitless
Capillary
- Range to
400 °C in 1 °C increments
- Back-pressure
design permits independent adjustment of split flow rate without
affecting column flow
- Septum purge
built in at 3 ml/min
- Accepts columns
up to 1.2-mm od
- Multimode
design includes split and splitless injection
- Splitless
purge time variable in 0.01-min elements
Electronic
Pressure Control
- Pressure
Control
- Programming:
Three ramps, 0.1-99 psi/minute
- Pressure
Range for inlets: 1-100 psi
- Flow
range for inlets: 0-450 ml/min
- Constant
Flow Mode (Standard): Inlet pressure is adjusted real time
to maintain constant volumetric flow at column outlet.
- Maximum
Channels: Two inlet channels to control column head pressure,
Four auxiliary channels to control split vent flow (Gas Saver
EPC), makeup gas, fuel gases, or external devices (general-purpose
EPC)
- Precision
and Accuracy
- Pressure
Setpoint Accuracy: +/- 0.1 psi
- Flow
Setpoint Accuracy: +/- 0.001 ml/min
- Average
Linear Velocity Accuracy: +/- 0.1 0.001 cm/sec.
Flame Ionization
Detector
- Range to
450 °C
- Grounded
jet and current limited design for operator safety
- Push-button
flame ignition
- Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in. id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in. id capillary
jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in. id capillary
jet
- Minimum detectable:
<5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic
range: <+/- 10% over a 10 7 range with 0.018-in.
id jet
- Conditions:
column flow 50 ml/min, 45 ml/min H2, 650 ml/min air,
propane sample
Column Oven
- Operating
range: 4 °C above ambient to 450 °C
- Setpoint
entry: 1 °C for temperatures, 0.1 °C for program rates
- Programming:
Rates 0.1 to 70 °C per min, 650 min maximum run time, Three
ramps with initial/final holds
GenTech Scientific, Used HP 5890 II Gas Chromatograph with EPC, FID,HP 5890/6890 GC Starter Kit, HP 5890 Series II GC with EPC and Dual FID
|