2790 Alliance
$19,900*
This
Waters 2790 Alliance is in excellent condition
and has been fully tested at GenTech Scientific. It
is an integrated solvent and sample management platform
which combines two traditional high-performance liquid chromatography
(HPLC) components: a solvent management system and a sample
management system. Includes NINETY
(90) DAY WARRANTY.
*All
prices are in U.S. dollars and do not include shipping,
handling and installation. International
Cooperating Organizations
Back
to Waters HPLC Systems
2790 Specifications
|
|

|
2790
Specifications:
|
Physical
Specifications
|
- Height:
22.5 in. (57 cm)
- Depth:
23.5 in. (60 cm)
- Width:
23 in. (58 cm)
- Weight:
136 lb. (62 kg)
- Wetted
Surface Materials: 316 stainless steel, UHMWPE, sapphire,
ruby, Tefzel® (ETFE), Teflon® (FEP and PTFE), PEEK,
Fluoroloy G, Fluoroloy–08R
|
|
Environmental
Specifications
|
- Operating
Temperature 4 to 40° C
- Relative
Humidity 20 to 80%, noncondensing
- Acoustic
Noise <65 dB(A)
|
|
Solvent
Management System Specifications
|
- Number
of Solvents: One to four, three with vacuum degas
- Solvent
Conditioning:
Vacuum degas, three solvents plus purge solvent
- Typical
Operating Flow Rate Range:
0.050 to 5.000 mL/min in 0.001-mL/min increments
-
Programmable Flow Rate Range:
0.000 and 0.010 to 10.000 mL/min in 0.001-mL/min increments
- Compressibility
Compensation:
Automatic and continuous
- System
Delay Volume: <400 µL, independent of backpressure,
1 mL/min
- Plunger
Seal Wash: Integral, active, programmable
- Gradient
Profiles: Eleven (11) gradient curves [including linear,
step (2 curves), concave (4 curves), and convex (4 curves)]
- Dry
Prime/Wet Prime:
Automatic, front panel control
- Flow
Ramping:
Time (0.01 to 30.00 min in 0.01-min increments) to reach
maximum flow rate
- Maximum
Operating Pressure: 5000 psi (345 bar) at 0.010 to 3.000
mL/min; Programmable upper and lower limits
- Composition
Range: 0.0 to 100.0% in 0.1% increments
- Composition
Accuracy: ±0.5% absolute, independent of backpressure
(Proportioning Valve Pair Test [degassed methanol: methanol/propylparaben,
2 mL/min, 257 nm])
- A
Composition Precision: <=0.15% RSD o r<=0.02 min
SD, whichever is greater, based on retention time. (Degassed
60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates,
phenone mix, 257 nm)
- Flow
Precision:
<=0.075% RSD or <=0.02 min SD, whichever is greater,
based on retention time (N = 6) or volumetric measures (0.200
to 5.000 mL/min), isocratic premix
- Flow
Accuracy: <=1% or 10 µL/min, whichever is greater,
(0.200 to 5.000 mL/ min), degassed methanol, at 600 psi
backpressure
|
|
Sample
Management System Specifications
|
- Number
of Sample Plates: Total of four plates:
- Microtiter plate – 96- or 384-well
- Vial plate – 2.0-mL vial (48)
- Tube plate – 0.5-mL microcentrifuge tube (48) or 1.5-mL
microcentrifuge tube (24)
- Open access plate – 2.0-mL vial (24 or 48)
- Sample
Temperature Control: 4 to 40° C, programmable in 1°
C increments
- Maximum
Sample Capacity: 1536
in four 384-well plates
- Number
of Sample Injections: 1 to 99 injections per sample
- Sample
Delivery Precision: <0.3% RSD, full loop 50 µL. (Default
wash/purge conditions, degassed 60% methanol: 40% water
Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
- Sample
Volume Linearity: >0.999% correlation, 5 to 20 µL, partial
fill in 50-µL loop, 3-µL pre- and post-sample air gaps (Default
wash/purge conditions, degassed 60% methanol:40% water Dial-a-Mix,
1 mL/min, six replicates, paraben mix, 257 nm)
- Needle
Wash Solvents:
Two: wash (strong solvent) and purge (sample- compatible)
solvent A
- Sample
Carryover: <0.01% or <2.5 nL (default wash/purge volumes),
whichever is greater. (Blank injection, 25 µL, 60% methanol:40%
water, after six replicates, 25 µL propylparaben, 1 g/L,
257 nm)
|
|
Instrument
Control Specifications
|
- Parallel
Sample Processing: User-programmable next-sample aspiration
and loop fill, column equilibration, just-in-time gradient
delivery
- Advanced
Operations: Auto addition, sample pooling, open access,
priority samples
- Column
Heater: 20° C (5° C above ambient) to 60° C, in 1° C
increments
|