NEW Multiple Gas Analyzer + Sulfur GC Systems
$20,995*
Many
analysts require natural gas analysis for BTU value calculations
or drilling and mudlogging applications. Frequently, sulfur
compounds are also of interest.
Because
low sulfur concentrations (<50ppm) are difficult to measure,
SRI has enhanced our popular Mulitple Gas Analyzer
GC systems to simultaneously monitor low levels
of sulfur compounds. The additional hardware required is
an FPD/FID detector, which selectively detects sulfur down
to mid-ppb range, and a room temperature Silcosteel sample
loop.
One
reason sulfur is so difficult to measure is that it disappears
on contact with hot stainless steel surfaces; even limited
contact with a hot stainless steel loop will destroy any
sulfur in the gas sample. Our solution is to use a Silcosteel-lined
transfer line leading to a splitter, and a 60-meter thick
film capillary column. While Teflon columns are sometimes
also used for sulfur analysis, the natural gas analysis
(using MoleSieve and Silica Gel) requires column temperatures
of 250oC or higher. Since the sulfur column is located in
the column oven, it is essential to use a column like the
60-meter capillary which can tolerate the higher temperatures.