"DISCOVER THE DIFFERENCE"

ANALYTICAL INSTRUMENT

SALES, SERVICE, PARTS, TRAINING

Waters 2790 Alliance

Listed Under Waters HPLC

Waters 2790 Alliance
$13,900*

This Waters 2790 Alliance is in excellent condition and comes fully tested at GenTech Scientific. It is an integrated solvent and sample management platform which combines two traditional high-performance liquid chromatography (HPLC) components: a solvent management system and a sample management system.

Includes NINETY (90) DAY WARRANTY.

*All prices are in U.S. dollars and do not include shipping, handling and installation.

Waters-2790-Alliance

Waters 2790 Alliance

2790 Specifications:

Physical Specifications
  • Height: 22.5 in. (57 cm)
  • Depth: 23.5 in. (60 cm)
  • Width: 23 in. (58 cm)
  • Weight: 136 lb. (62 kg)
  • Wetted Surface Materials: 316 stainless steel, UHMWPE, sapphire, ruby, Tefzel® (ETFE), Teflon® (FEP and PTFE), PEEK, Fluoroloy G, Fluoroloy–08R
Environmental Specifications
  • Operating Temperature 4 to 40° C
  • Relative Humidity 20 to 80%, noncondensing
  • Acoustic Noise <65 dB(A)
Solvent Management System Specifications
  • Number of Solvents: One to four, three with vacuum degas
  • Solvent Conditioning: Vacuum degas, three solvents plus purge solvent
  • Typical Operating Flow Rate Range: 0.050 to 5.000 mL/min in 0.001-mL/min increments
  • Programmable Flow Rate Range: 0.000 and 0.010 to 10.000 mL/min in 0.001-mL/min increments
  • Compressibility Compensation: Automatic and continuous
  • System Delay Volume: <400 µL, independent of backpressure, 1 mL/min
  • Plunger Seal Wash: Integral, active, programmable
  • Gradient Profiles: Eleven (11) gradient curves [including linear, step (2 curves), concave (4 curves), and convex (4 curves)]
  • Dry Prime/Wet Prime: Automatic, front panel control
  • Flow Ramping: Time (0.01 to 30.00 min in 0.01-min increments) to reach maximum flow rate
  • Maximum Operating Pressure: 5000 psi (345 bar) at 0.010 to 3.000 mL/min; Programmable upper and lower limits
  • Composition Range: 0.0 to 100.0% in 0.1% increments
  • Composition Accuracy: ±0.5% absolute, independent of backpressure (Proportioning Valve Pair Test [degassed methanol: methanol/propylparaben, 2 mL/min, 257 nm])
  • A Composition Precision: <=0.15% RSD o r<=0.02 min SD, whichever is greater, based on retention time. (Degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, phenone mix, 257 nm)
  • Flow Precision: <=0.075% RSD or <=0.02 min SD, whichever is greater, based on retention time (N = 6) or volumetric measures (0.200 to 5.000 mL/min), isocratic premix
  • Flow Accuracy: <=1% or 10 µL/min, whichever is greater, (0.200 to 5.000 mL/ min), degassed methanol, at 600 psi backpressure
Sample Management System Specifications
  • Number of Sample Plates: Total of four plates:
    - Microtiter plate – 96- or 384-well
    - Vial plate – 2.0-mL vial (48)
    - Tube plate – 0.5-mL microcentrifuge tube (48) or 1.5-mL microcentrifuge tube (24)
    - Open access plate – 2.0-mL vial (24 or 48)
  • Sample Temperature Control: 4 to 40° C, programmable in 1° C increments
  • Maximum Sample Capacity: 1536 in four 384-well plates
  • Number of Sample Injections: 1 to 99 injections per sample
  • Sample Delivery Precision: <0.3% RSD, full loop 50 µL. (Default wash/purge conditions, degassed 60% methanol: 40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
  • Sample Volume Linearity: >0.999% correlation, 5 to 20 µL, partial fill in 50-µL loop, 3-µL pre- and post-sample air gaps (Default wash/purge conditions, degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
  • Needle Wash Solvents: Two: wash (strong solvent) and purge (sample- compatible) solvent A
  • Sample Carryover: <0.01% or <2.5 nL (default wash/purge volumes), whichever is greater. (Blank injection, 25 µL, 60% methanol:40% water, after six replicates, 25 µL propylparaben, 1 g/L, 257 nm)
Instrument Control Specifications
  • Parallel Sample Processing: User-programmable next-sample aspiration and loop fill, column equilibration, just-in-time gradient delivery
  • Advanced Operations: Auto addition, sample pooling, open access, priority samples
  • Column Heater: 20° C (5° C above ambient) to 60° C, in 1° C increments