6890A GC with Dual Split/Splitless Injector/Dual FID
$24,000
Agilent/HP
6890 GC System includes dual Split/Splitless Injectors,
dual Flame Ionization Detector, Electronic Pressure
Control on detector and injectors, HPIB Communications
installed on GC, manuals, and ONE
(1) YEAR WARRANTY.
Additional
or alternative detectors available.
HP
5890/6890 GC Starter Kit - $978
*All
prices are in U.S. dollars and do not include shipping, handling
and installation. International
Cooperating Organizations
|
| |
| Specifications
Data
communications
- HP-IB,
RS-232-C, and two analog output channels (1-mV, 1-V, and
10-V output available) as standard
- INET
interface optional
Heated
zones
- Independent
heated zones, not including oven: six (two inlets, two detectors,
and two auxiliary)
- Maximum operating
temperatures for auxiliary zones: 400° C
Column
oven
- Operating
temperature: 4° C above ambient to 450° C
- Temperature
setpoint resolution: 1° C
- Maximum
run time: 999.99 min
- Programming
ramps/plateaus: 6/7
Inlets
- All
inlets include septum purge.
- Split/splitless:
electronic flow, 400° C max, electronic entry of pressure
or flow and split ratio version.
- Pressure
setting range: 0-100 psi
- Total
flow setting range: split/splitless, 0-200 ml/min N2,
0-1000 ml/min H2 or He
Detectors:
All detectors include electronic on/off of
all detector gases.
FID
- 450°
C maximum operating temperature
- Automatic
flame ignition from the keypad or ChemStation
- Flame
out detection
- MDL:
<5 pg carbon/s as propane using N2 carrier and
0.2794-mm jet
- Linear
dynamic range: < 10%, 107 with N2
carrier and 0.2794-mm jet
NPD
- 400°
C maximum operating temperature
- Automatic
baseline adjusting via keypad or ChemStation
- MDL:
<0.4 pg N/s, <0.2 pg P/s with azobenzene/malathion mixture
- Selectivity:
35000 to 1 gN/gC, 75000 to 1 gP/gC with azobenzene/malathion/
octadecane
- Dynamic
range: >105 N >105 P with azobenzene/malathion
mixture
TCD
- 400°
C maximum operating temperature
- Single
filament with microcell
- MDL:
<400 pg propane/ml He caxrier (MDL may be affected by laboratory
environment.)
- Linear
dynamic range: 105 (5%)
ECD
- Equipped
with anode purge for contamination resistance
- 400°
C maximum operating temperature
- Makeup
gas types: argon (5%)/methane, nitrogen
- Radioactive
source: 15 mCi 63Ni plated
- MDL:
<0.04 pg/s lindane
- Dynamic
range: > 104 with lindane