Waters 717 Plus HPLC Autosampler, 486 Tunable Absorbance Detector,
600 Pump, and 600 Controller
$11,100*

The Waters 717 Plus Autosampler injects samples into high performance liquid chromatography system. The Waters 717 Plus automates sampling by storing up to 96 sampling methods or "step". Each method controls all injection parameters for a sequence of vials. Includes 48-position tray for standard sized vials, and a 96-position microvial tray is available if needed.

The Waters 486 Tunable Absorbance Detector is a single-channel tunable ultraviolet/visible detector designed for high performance liquid chromatography applications. The Waters 486 can operate as a stand-alone unit or with other Waters system modules. The Waters 486 provides a usable light intensity over the wavelength range of 190-600 nm.

Waters 600 Pump with stainless steel flow path features 50 ul pump head volume, maximum pressure 5000 psi (351.5 kg/cm2) and 0.01 to 5.0 ml/min operating flow rate.

Waters 600 Controller features LCD Display; program tables: flow rate, gradient, timed events 15 time-based step/table, 15 table capacity; and time-based multi-method capacity. Includes linking program table.

Includes: 96 Position microvial tray.

All above instruments are in good working condition and tested at GenTech. Include NINETY (90) DAY WARRANTY.

SRI Model 203 Single Channel Chromatography Data System** includes PeakSimple Software to run with Windows 95, 98 Me, XP, and Windows NT 4.0, allows acquisition, integration and display of one detector channel from a GC or LC. Includes two independent timebases, two 0-5VDC DACs, two remote start inputs and built-in AC power supply. Also includes 24-bit serially interfaced acquisition hardware, manuals, FREE lifetime software upgrades and TWO (2) YEAR WARRANTY.

**GenTech is a distributor of new SRI instruments Does not include computer - requires Win 95 or newer.

*All prices are in U.S. dollars and do not include shipping, handling and installation. International Cooperating Organizations

Back to Waters HPLC Systems
717 Specifications | 486 Specifications

request quote on used laboratory equipment

Waters 717 Plus HPLC Autosampler, 486 Tunable Absorbance Detector,  616 Pump & 600s Controller

717 Specifications:

Operational Specifications
  • Injection Volume Range: 0.1 uL to 2000 uL in 0.1 uL Increments
  • Linearity (1 to 100 uL): > 0.99 Coefficient of deviation
  • Carryover: < 0.1 %
  • Bandspreading: < 10% Contribution for all solvent systems
  • Precision (typical, 5 to 50 uL): <0.5 % RSD
  • System Operating Pressure: 4500 psig maximum continuous operating pressure 5000 psig maximum surge for <30 seconds
  • Wash Flow, Solvent Rate: > 0.4 mL per injection
  • Minimum Sample Required: 7 uL using low volume inserts
Mechanical Specifications
  • Wetted Surface Materials: 316 Stainless Steel, Teflon, Glass, Fluoraloy-08 R, UHMWPE
  • Number of Sample Vials: 1 to 96
  • Number of Injections: Programmable 1 to 99 injections per vial
  • Allowable Solvents: All standard LC solvents
  • Inlet Tubing to Injector: 0.040 I.D. stainless steel
  • Outlet Tubing from Injector: 0.009 ID stainless steel
  • Total Solvent Volume: 836 uL
Environmental Specifications
  • Operating Temperature Range: 4-40 °C
  • Relative Humidity: 80%, non-condensing
  • Storage Temperature Range: - 40 to 70 °C

Back to top


486 Specifications:

Operational Specifications
  • Wavelength range (single, fixed wavelength): 190 to 600 nm
  • Wavelength accuracy: ± 2 nm
  • Wavelength repeatability: ± 0.25 nm
  • Noise: ± 1.0 x 10-5 AU, 1 second filter setting, 210 to 280 nm) dry cell
  • Drift:
    - 1.0 x 10-4 AU per °C (after warm-up)
    - 1.0 x 10-4 AU per hour (after warm-up)
  • Sensitivity Setting Range: 0.001 to 2.0 AUFS (under software control)
  • Filter Setting Range: 0.1 to 5.0 seconds (under software control)
  • Bandwidth: 8 nm
Optical Component Specifications
  • Lamp Source: Deuterium arc lamp
  • Flow cell: Dual beam Taper-Cell design
  • Path length: 10 mm (standard, analytical)
  • Cell volume: 8 ul (standard, analytical)
  • Pressure limit: 1000 psi
  • Materials: 316 stainless steel, Dynasil Suprasil II, Teflon
Environmental Specifications
  • Operating temperature: 15 to 32.2 °C
  • Operating humidity: 20 to 80%, non-condensing
  • Shipping and storage temperature: -40° to 65 °C
  • Shipping and storage humidity: 20 to 80% non-condensing

Terms & Conditions