| Number
of flow lines |
DGU-20A5:
5 |
|
Degassing performance |
UV
absorption difference with and without degasser is
over 300 mAU (using UV absorption detector, wavelength
210 nm, sensitivity 0.5 AU, flowing Methanol saturated
with dissolved Oxygen (HPLC grade) at 25°C with 1
ml/min) |
|
Max. operating
flow rate |
20
mL/min per flow line (using deionised water – applicable
range varies depending on the solvent being delivered). |
| Internal
capacity |
380
μL per flow line |
|
Pressure resistance |
98
kPa (1 kgf/cm2) or less (Only when suctioning with
the delivery unit. Pressurized delivery to the degasser
is prohibited.) |
| Tubing
|
3
mm O.D. x 1.5 mm I.D. PTFE tubing |
|
Materials contacting
solvents |
Special
synthetic polymers designed for all solvents (degassing
membrane), PEEK (connector joints) |
|
External output |
When
system self-diagnostics detects an abnormality, a
signal is output to the external terminal (ALARM) |
|
Self-cleaning |
Includes
mechanism that extends vacuum pump life by drawing
in air when the vacuum pump is running |
| Dimensions
and weight |
W
260 x H 72 x D 421 mm; 5 kg |
| Operating
conditions |
4
– 35°C, 20 – 85% humidity |
| Power
supply |
+24
V (powered via connection by remote cable to delivery
unit) |