Quant
II lets you run up to 10 standards for quantitative
analysis
-
Quant
I lets you run wavelength scanning with auto scaling,
comparison, and customizing
-
Auto-6
lets you run up to 6 samples for kinetic analysis, and
supports custom begin and end parameters, and the Lineweaver-Burke
plot
-
Gel-Scan
for tube gel and film studies
-
Five
Carat cell for microsampling
SPECIFICATION
Operating
Principle: Double beam spectrometer for the UV/VIS
region, with microcomputer electronics, keyboard entry and
vacuum fluorescence display.
Monochromator:
Holographic concave grating with 1053 lines/mm.
Radiation:
Prealigned Deuterium and Tungsten halogen lamps with automatic
lamp switching.
Wavelength:
190 nm – 1100 nm.
Stray Radiation:
Measured with a 2 sec response time
< 0.03 %T at 220 nm
< 0.02 %T at 340 nm, 370 nm.
Wavelength
Accuracy: +- 0.3 nm at Deuterium peak 656.1 nm.
Reproducibility:
+- 0.1 nm.
Spectral
Slit Width: 2 nm.
Photometric
Accuracy: +- 0.005 A at 1 A (measured with NBS 930
filters).
Reproducibility:
+- 0.002 A at 1A.
Range:
-3 to +3 A.
Stability:
< 0.0003 A/h (at 500 nm, 1 sec response time, after warmup).
Baseline
Stability: +- 0.001 A (200-1100 nm at 0 A, 2 sec
response time, 240 nm/min scan speed.
Noise:
< 0.0003 A (at 0 A, 500 nm, 2 sec response time)
|