5890 Series II GC with Dual FID
$7,995*
Includes
Dual Split/Splitless Injectors, Dual Flame Ionization
Detectors, INET Communications installed on GC,
cables, manuals and ONE (1) YEAR
WARRANTY. Other injectors and detectors available.
HPIB Communications and Data System available.
Available
in HP Parts.
HP
5890/6890 GC Starter Kit - $978
*All
prices are in U.S. dollars and do not include shipping, handling
and installation. International
Cooperating Organizations
|
|
|
Specifications:
Heated Devices
- Five heated
zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored:
Two
Split/Splitless
Capillary
- Range to
400° C in 1° C increments
- Back-pressure
design permits independent adjustment of split flow rate without
affecting column flow.
- Septum purge
built in at 3 ml/min
- Accepts columns
up to 1.2-mm od
- Multimode
design includes split and splitless injection.
- Splitless
purge time variable in 0.01-min elements
Flame Ionization
Detector
- Range to
450° C
- Grounded
jet and current limited design for operator safety
- Push-button
flame ignition
- Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in. id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in. id capillary
jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in. id capillary
jet
- Minimum detectable:
<5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic
range: <+/- 10% over a 10 7 range with 0.018-in.
id jet
- Conditions:
column flow 50 ml/min, 45 ml/min H2, 650 ml/min air,
propane sample
Column Oven
- Operating
range: 4° C above ambient to 450° C
- Setpoint
entry: 1° C for temperatures, 0.1° C for program rates
- Programming:
Rates 0.1 to 70° C per min, 650 min maximum run time, three
ramps with initial/final holds